4
Instructions for Use
General Information
The complete pumping system consists of the VacIon pump, its
magnet, and a power supply. These VacIon pump is a pump which
operates on the principle of ion gettering in the pressure range from
-3
10
the system pressure from atmosphere to below 10
The pump should be powered by a positive voltage of 3000 – 3500
Volt (open circuit), direct current supply and the short circuit
current of the control unit should match the current drawn by the
pump at its maximum operating pressure.
40/52
Torr to less than 10
Miniature and 2 l/s VacIon Pumps User Manual 87-400-044-01 (E)
-9
Torr. A roughing pump is needed to lower
-3
Torr.