Behavior
during use
Pacing with
high rates
Behavior after use
Battery operation
1)
Reocor S reminds you to replace the battery with the ERI signal (Low Battery LED flashes).
During use of Reocor S, the protective cover must be
completely closed to prevent accidental resetting of the
programmed parameters.
Secure Reocor S either horizontally on a non-slip surface
or on the patient with an armband, or operate it from a
hanging position on the infusion stand using the hanger
on the back of the device.
Reocor S must not be worn directly on the skin.
During use of Reocor S, the heart rate of the patient must
be monitored with an ECG monitor with alarm function.
In case of disturbances caused by electromagnetic inter-
ference (EMI), Reocor S will switch to operating mode
S00 when certain limits are exceeded.
Pacing the heart with rates higher than 180 ppm over a
long period of time can cause severe hemodynamic com-
plications. Pacing with high rates should only be per-
formed when continuous monitoring is ensured.
After a defibrillation or cauterization, the device should
be subjected to a function test.
If the device will be stored for a long period of time, the
battery should be removed to prevent damage due to
leakage.
A damp cloth and mild soap can be used for cleaning.
Strong cleaning agents or organic solvents should be
avoided, as these can corrode the plastic housing.
Inspection and maintenance work should be performed
according to page 25.
Do not use rechargeable batteries. The service time of
these batteries is difficult to estimate, making it possible
to inadvertently exceed the ERI
den cessation of pacing.
Only 9-volt batteries with the international code
IEC 6LR61 must be used. When using the battery type
MN 1604 Duracell
for at least 600 hours before the battery must be
replaced.
It is possible to exchange a battery while Reocor S is in
use. The device remains ready for use for at least 30 s at
the ambient temperature (20 ± 2°C) when the battery is
removed.
For safety reasons, the patient should be paced by
another source during the battery replacement.
1)
time, resulting in sud-
®
®
Procell
, external pacing is possible