TP 37 / TP 3M
Fuse
The fuses of the calibrator are located on the front and are integrated into the mains connec-
tion. If the display remains dark when the mains voltage is applied and the fan does not run,
check the fuses and replace them if necessary.
Pull the mains connection cable from the calibrator.
Prise open the fuse compartment from the bottom with a fingernail or a flat screwdriver.
Remove the compartment with the fuses.
Check the fuses and replace both fuses.
Only use fuses of the same type:
Type
37200E.2(i)
T6.3AH
Fuse
250V
Fit the fuse compartment back in place and connect the mains connecting cable
Should the fuses blow repeatedly, there is probably a fault in the calibrator. In this case, send
the calibrator to SIKA for repairs ( p. 57).
Technical changes reserved
TP
TP
37165E.2(i)
T6.3AH
250V
TP
TP
37700E.2(i)
37450E.2(i)
T10AH 250V T10AH 250V
Maintenance and Cleaning
TP
TP
3M165E.2(i)
3M255E.2(i)
T6.3AH
T6.3AH
250V
250V