TP 37 / TP 3M
Fuse
The fuses of the calibrator are located on the front and are integrated into the mains connec-
tion. If the display remains dark when the mains voltage is applied and the fan does not run,
check the fuses and replace them if necessary.
 Pull the mains connection cable from the calibrator.
 Prise open the fuse compartment from the bottom with a fingernail or a flat screwdriver.
 Remove the compartment with the fuses.
 Check the fuses and replace both fuses.
Only use fuses of the same type:
Type
37200E.2(i)
T6.3AH
Fuse
250V
 Fit the fuse compartment back in place and connect the mains connecting cable
Should the fuses blow repeatedly, there is probably a fault in the calibrator. In this case, send
the calibrator to SIKA for repairs ( p. 57).
Technical changes reserved
TP
TP
37165E.2(i)
T6.3AH
250V
TP
TP
37700E.2(i)
37450E.2(i)
T10AH 250V T10AH 250V
Maintenance and Cleaning
TP
TP
3M165E.2(i)
3M255E.2(i)
T6.3AH
T6.3AH
250V
250V
- 59 -