Commissioning and Operation
4.2 Electrical Connection
The voltage specified on the type plate must match the mains voltage.
Only connect the calibrator to a properly installed and earthed 3-pin socket suitable for
earthed plugs.
TP 37700E.2(i) and TP 37450E.2(i): Protective conductor (PE) must be available.
Do not use any extension cables or adapter plugs.
IMPORTANT
As a replacement, only use original SIKA cables or cables of the same construc-
tion with the corresponding specification (H05VV-F 3 G 0.75 mm² with angled pro-
tective contact plugs and cold appliance socket, length approx. 2 m).
Connect the mains cable to the calibrator's appliance plug.
Insert the plug of the mains cable into a suitable socket.