Commissioning and Operation
4.3.4 Calibration Bath
TP
Type
37200E.2(i)
Calibration
-/-
bath
WARNING
When using the calibration bath, calibration liquid may splash out.
 Always wear safety goggles when handling calibration liquids.
The calibration bath is used for calibrating sensors with special shapes or dimensions. Direct
contact between the sensor and the calibration liquid ensures excellent heat transfer. The
calibration liquid is poured directly into the tank or into a tub insert.
The calibration bath includes the transport lid, the working lid, the sensor cage, the magnetic
stirrer, the drain syringe, the magnetic lifter and as an accessory, the tub insert.
Transportation lid
The transport cover is used to securely close the calibration bath. It prevents the calibration
liquid from leaking during transport.
CAUTION
The safety valve of the transport lid triggers at a pressure of
~1.5 bar. Hot vapours may escape.
 Always unscrew the transport lid before operating the calibra-
tion bath.
 Do not screw the transportation lid back on until the calibration
bath has cooled down.
Working lid
The working lid fulfils various tasks during the operation.
It reduces the evaporation of the calibration liquid to a minimum.
It reduces the cooling on the surface of the calibration liquid.
It ensures stable positioning of the DUTs in the calibration bath.
The working lid is screwed on to the calibration bath and has five openings
for DUTs. The unused openings can be closed with suitable silicone plugs.
- 42 -
TP
TP
37165E.2(i)
37700E.2(i)
-/-
-/-
TP
TP
37450E.2(i)
3M165E.2(i)
-/-
© SIKA • Ba_TP37-TP3M • 10/2023
TP 37 / TP 3M
TP
3M255E.2(i)