TP 37 / TP 3M
4
Commissioning and Operation
4.1 Operating Conditions
Installation site and operating position
Only suitable for indoor use, do not use outdoors.
Only operate vertically on a level surface. The surface must be firm, clean and dry.
The stability and the specified properties of the calibrator cannot be guaranteed if the cali-
brator is set up differently.
At higher test temperatures, use a sufficiently large, fireproof base.
Sufficient clearance around the device.
To the front > 1 m, to the sides and to the rear > 0.5 m. Sufficient free space above the
device.
The device under test must be securely fixed in the calibrator.
Ensure sufficient ventilation.
Do not use in potentially explosive atmospheres (flammable or explosive atmospheres).
Do not operate near flammable materials.
Do not install in a cabinet or other enclosure.
Do not block or cover ventilation openings.
Place the device in such a way that it can be switched off at any time.
Series
Ambient conditions
Operating temperature
Transport and storage temperature
Humidity (RH)
Electrical safety
Operating conditions
Location
Altitude
Operating position
Application area
IMPORTANT
The plug of the mains connecting cable serves as a "KILL" switch.
The plug must always be freely accessible and easily reachable.
In an emergency, pull the plug to disconnect the device from the mains.
Technical changes reserved
< 80% to 31 °C, reducing linearly up to 50% at 40 °C
(non-condensing environment)
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree
2 according to IEC-61010-1:2001
Standing upright/vertically
tested according to EN 61326-1, class A (industrial sector)
Commissioning and Operation
TP 37 / TP 3M
0...50 °C
-10...60 °C
Interiors
Up to 2000 m
- 37 -