TP 37 / TP 3M
4
Commissioning and Operation
4.1 Operating Conditions
Installation site and operating position
 Only suitable for indoor use, do not use outdoors.
 Only operate vertically on a level surface. The surface must be firm, clean and dry.
The stability and the specified properties of the calibrator cannot be guaranteed if the cali-
brator is set up differently.
 At higher test temperatures, use a sufficiently large, fireproof base.
 Sufficient clearance around the device.
To the front > 1 m, to the sides and to the rear > 0.5 m. Sufficient free space above the
device.
 The device under test must be securely fixed in the calibrator.
 Ensure sufficient ventilation.
 Do not use in potentially explosive atmospheres (flammable or explosive atmospheres).
 Do not operate near flammable materials.
 Do not install in a cabinet or other enclosure.
 Do not block or cover ventilation openings.
 Place the device in such a way that it can be switched off at any time.
Series
Ambient conditions
Operating temperature
Transport and storage temperature
Humidity (RH)
Electrical safety
Operating conditions
 Location
 Altitude
 Operating position
Application area
IMPORTANT
The plug of the mains connecting cable serves as a "KILL" switch.
 The plug must always be freely accessible and easily reachable.
 In an emergency, pull the plug to disconnect the device from the mains.
Technical changes reserved
< 80% to 31 °C, reducing linearly up to 50% at 40 °C
(non-condensing environment)
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree
2 according to IEC-61010-1:2001
Standing upright/vertically
tested according to EN 61326-1, class A (industrial sector)
Commissioning and Operation
TP 37 / TP 3M
0...50 °C
-10...60 °C
Interiors
Up to 2000 m
- 37 -