4.3 Preparing the Calibrator
4.3.1 Dry Block Calibration
Type
37200E.2(i)
Dry block
Calibration inserts with single or multiple holes are used for the calibration of straight temper-
ature sensors.
To achieve the accuracy of the calibrators, the temperature sensor (DUT) and
the calibration insert must be matched to one another:
The borehole of the calibration insert may be a maximum of 0.5 mm larger
than the diameter of the test specimen.
The measurement element of the DUT must be located in the homogeneous
temperature zone of the calibration insert.
Inserting
Only use the supplied calibration inserts
made of the appropriate material.
If in doubt, please contact SIKA.
 Insert the appropriate calibration insert
into the metal block with the exchange
tool.
 Only Air Shield Insert:
Push the calibration insert down to the
stop.
 Center the calibration insert so that there
is an even air gap between the calibration
insert and the block.
External reference sensor (optional):
 Align the calibration insert so that the hole for the external reference sensor is located at
12 o'clock.