TP 37 / TP 3M
6
Calibration
WARM UP BEFORE THE FIRST CALIBRATION
Allow the calibrator to warm up for at least one hour before the first calibration,
e.g. by running the calibrator to the first test point.
IMPORTANT
Before calibrating, check that the inside of the block and the surface of the cali-
bration insert are dry. Any ice or condensation that is present can be safely re-
moved by heating to above 100 °C.
6.1 Calibration Setup
If you do not need a certificate after calibration, please continue with "Calibration Without
Certificate" ( p. 55).
For a calibration with certificate, you must first add your devices under test (temperature sen-
sors) and a test task. A test task contains the test points, one or more devices under test and
the calibration function used.
From the main screen, tap
You can now set up test tasks, devices under test and calibration functions.
Technical changes reserved
Setup Calibration.
Calibration