Declaration Of Conformity; Application; Technical Characteristics; Input - PR electronics 5133 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

DECLARATION OF CONFORMITY

As manufacturer
PR electronics A/S
Lerbakken 10
DK-8410 Rønde
hereby declares that the following product:
Type: 5133
Name: 2-wire RTD transmitter
is in conformity with the following directives and standards:
EMC directive 89/336/EEC and later amendments
From serial no.: 960194 ff
EN 61 326
EN 50 081-1, EN 50 081-2
EN 50 082-1, EN 50 082-2
This declaration is issued in compliance with article 10, subclause 1 of the
EMC directive. For specification of the acceptable EMC performance level,
refer to the electrical specifications for the module.
Rønde, 1 June 1996
10
Peter Rasmussen
Manufacturer's signature
2-WIRE RTD TRANSMITTER 5133
Programmable via PC
Input for Pt100, Ni100, or Ohm
Linearised 4...20 mA analogue output
2-wire connection
1- or 2-channel version
DIN rail mounting

APPLICATION:

Linearised temperature measurement with Pt100 or Ni100 sensor. Conversion of
linear resistance change to standard analogue current / voltage signal from e.g.
valves or linear movements with attached potentiometer.
4...20 mA signal simulator via externally mounted 10-turn potentiometer.

TECHNICAL CHARACTERISTICS:

General:
The transmitter may be configured to the requested signal type by means of a
DOS-based PC with the configuration program PReset 5000 and the communi-
cations unit Loop Link 5905A. The configuration of the transmitter may be done
without connection of supply voltage as the communications unit supplies the
required power. The supply voltage of 10...35 VDC is converted to a 4...20 mA
signal in the 2-wire connection. The output has voltage supply ground as refe-
rence and is protected against polarity reversal. The sensor connection is always
a 3-wire connection with cable compensation for up to 50 Ω in each wire. If a
2-wire connection is requested, pins 43 and 42 (chan nel 1) or 53 and 52 (chan-
nel 2) must be short-circuited in the connection plug (no cable compensation).
Sensor error detection is as standard set to Upscale, Downscale or entered to
special output value.

INPUT:

Linearised Pt100 temperature input according to IEC 751 within the range
-200...+850°C. Linearised Ni100 temperature input according to DIN 43760 within
the range -50...+250°C. Linear resistance input within the range 0...10 kΩ. Min.
span for the RTD input is 50°C, for linear re sistance 30 Ω.