High-level disinfection of the mirror sleeve (without HLD Set)
8.1.3.9
In case the HLD Set is not available
NOTICE
Dry heat sterilization and high-level disinfection must not be combined.
1. Pre-cleaning (see „Pre-cleaning of the mirror sleeve [ → 59]").
2. Remove the mirror sleeve from the CEREC Omnicam (see
„Removing the mirror sleeve [ → 58]").
3. Place the camera in its camera cradle.
4. Use the following disinfectant for the high-level disinfection: CIDEX
OPA.
CAUTION
Observe the disinfectant's manufacturer's safety indications!
NOTICE
Only place the mirror sleeve into the container and not the entire
CEREC Omnicam.
Ensure that the mirror sleeve is placed upright in the disinfectant (with
the coated sapphire glass down) so that the solution cannot penetrate
the inside of the mirror sleeve.
5. Choose a container. Find an example on the left.