6 Installation and startup
6.5 Commissioning
32
Operating Instructions (valid for Canada) CEREC AF, CEREC AF Connect
Dimensions of the PC as of hardware version 3.4.2
Width
in mm
180
in inches
7
A
PC
B
Cabinet
C
Cabinet fan
NOTICE
The temperature inside the cabinet affects the performance of the PC.
Full performance can only be guaranteed up to a temperature of 35°C
(95°F) (measured at the air intake of the PC). A drop in performance can
become apparent if the scanning process slows down when the
temperature rises above 35°C (95°F).
- The cabinet should have an active ventilation or air extraction system;
see the sketch for an example.
- It is particularly important to check the temperature inside the cabinet
when the ambient temperature is higher (in summer, for example). If
necessary, open the door or place the PC somewhere else.
Dentsply Sirona
Depth
Height
360
435
14
17 ⅛
D
Ventilation slots in the cabinet
E
PC power supply unit/PC air
extraction
F
PC air intake
66 37 487 D3652
D3652.201.01.14.02 10.2017