Section 9
Each lab should use its own standards and procedures for performance.
Test known specimens/controls at each of the following events in
accordance with local, state, and/or federal regulations or accreditation
requirements.
•
Each new day of testing
•
A new canister of strips is opened
•
A new operator uses the analyzer
•
Test results seem inaccurate
•
After performing maintenance or service on the analyzer
If QC tests do not provide expected results, perform the following checks:
•
Ensure the strips used are not past their expiration date.
•
Ensure strips are fresh from a new canister.
•
Ensure the controls are not past their expiration date.
•
Repeat the test to ensure no errors were made during the test.
For US customers, call customer service toll free at 1-(800)-838-9502 for
additional information. For customers outside the US, contact your local
distributor.
Quality Control
36