Section 9
Each lab should use its own standards and procedures for performance.
Test known specimens/controls at each of the following events in
accordance with local, state, and/or federal regulations or accreditation
requirements.
Each new day of testing
A new canister of strips is opened
A new operator uses the analyzer
Test results seem inaccurate
After performing maintenance or service on the analyzer
If QC tests do not provide expected results, perform the following checks:
Ensure the strips used are not past their expiration date.
Ensure strips are fresh from a new canister.
Ensure the controls are not past their expiration date.
Repeat the test to ensure no errors were made during the test.
For US customers, call customer service toll free at 1-(800)-838-9502 for
additional information. For customers outside the US, contact your local
distributor.
Quality Control
36