Insert a control strip into the strip channel in the same direction as the arrows
indicated on the strip. Ensure that the test strip is inserted all the way.
Press
to start the optical check. If the meter displays YES, the meter is
normal. If the meter displays no, the meter is not functioning properly.
If the meter displays no, check the control strip for contamination or if it is
bent or damaged. If there are any visible signs of damage or contamination,
discard the control strip and retest using a new control strip.
Note: For US customers, call customer service toll free at
1-(800)-838-9502 if the meter displays no again. For
customers outside the US, contact your local distributor to
double check if any Problem with system.