InPro 8300 RAMS Optical Product Monitor
4.4 Electrical Connection of the InPro 8300 RAMS
The InPro 8300 RAMS is supplied with the connection cable firmly connected between active and pas-
sive part. Once the InPro 8300 RAMS has been installed mechanically, the electrical connections can
be established.
Procedure:
• Unscrew the window plate from the InPro 8300 RAMS
• Pass the stripped cable through the cable gland
– Cable diameter 4 ... 10 mm
– Core cross section 0.75 mm
• Connect the cable cores in accordance with the circuit diagram.
• Insert the desiccant capsule in the terminal compartment
• Replace the window plate
Tools required:
•
Open-ended wrench a/f 22 mm (not included)
•
Cable stripper (not included)
•
Screwdriver 2.4 mm (included in the InPro 8300 RAMS accessories kit)
•
Desiccant capsules (2 pcs, included in the InPro 8300 RAMS accessories kit)
4.4.1 Electrical connection of the InPro 8300 RAMS without connection module
RX
TX
Iin
1 – 4 = digital inputs / outputs (in analog mode inputs to the InPro 8300 RAMS in digital mode
Iout
L+
L-
© 04/09 Mettler-Toledo AG, Greifensee
Printed in Switzerland
= receiving line of the serial interface
= sending line of the serial interface
= external mA input (+)
outputs of the InPro 8300 RAMS)
= mA output (+)
= power supply, 24V DC
= ground
2
InPro 8300 RAMS Optical Product Monitor
13
52 801 115