If you wish to switch to digital mode, you can do so by clicking the "Change-over" button in the "Ope-
rating mode" frame and confirming your selection by clicking "Yes" in the message box that opens. If
you click "No" in the message box, the system remains in analog mode.

5.2.4 Zero adjustment

Carrying out zero adjustment
When you click this button, the program attempts to set the zero points for the InPro 8300 RAMS pa-
rameters. For the exact procedure and the points to be noted please refer to section 5.5.

5.2.5 Other settings

Moisture alarm
Relative humidity is measured in the InPro 8300 RAMS electronics. This value should always be be-
low 15 % as otherwise there is a danger of condensation forming. If the moisture alarm function is
active, an alarm will be triggered if a relative humidity of 15 % is overshot by reducing the mA output
to a value of 2.96 for approx. 1 second. This will cause a fault in the downstream control system which
can be interpreted as a moisture alarm. After replacing the desiccant capsules, the alarm must be ack-
nowledged by clicking "Reset".
Measuring point
In the "Measuring point" field a measuring point designation can be entered.
Created by
The name of the author of the project can be entered here.
Created on
This field cannot be accessed. The date of creation is set automatically (system date).