1.
Ensure the equipment is clean according to the previous steps.
2.
Prepare and condition the disinfection solution according to the solution
manufacturer's instructions and the conditions stated in
3.
Disinfect the video baton according to the conditions stated in
exposure process and times vary depending on the solution and the component.
IMPORTANT
component as often as needed to keep it visibly wet throughout the specified
exposure time. You may use multiple wipes if necessary.
4.
If applicable, rinse the video baton according to the solution manufacturer's
instructions.
5.
Dry the video baton as appropriate for the disinfectant you are using.
If you are using ...
A liquid disinfectant
Pre‑moistened wipes
6.
Inspect the video baton according to the instructions in the following procedure,
and then store the disinfected video baton in a clean environment.
PROCEDURE 5. CLEAN THE CHARGING CRADLE
Clean the charging cradle when it is visibly soiled and on a regular basis, as per a schedule
established by the medical care facility or provider.
CAUTION
1.
Ensure the monitor is not in the charging cradle, and then unplug the device.
2.
Using a solution from
until all visible contamination has been removed.
18
Disinfection
Level
Standard cycles in the following processors:
STERIS
Sterilization
SYSTEM 1E
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
Sterilization
Up to 200 cycles as per manufacturer's instructions
STERRAD
STERRAD
Sterilization
STERRAD
STERRAD
If you are using pre‑moistened wipes, wipe down the
Then ...
Use a sterile cloth, hospital‑grade clean air, or a
low‑temperature dryer to dry the component.
Allow the component to air dry completely.
Do not submerge the charging cradle in a liquid solution.