Data Interface - KERN AGB Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Data Interface

Purpose
Your balance comes equipped with an
interface port for connection to a computer
or other peripheral device. You can use an
on-line computer to change, start and/or
monitor the functions of the balance and
the application programs.
Female interface connector
25
13
Pin Assignment Chart, 25-pin
female interface connector, RS-232:
Pin 1: Shield
Pin 2: Data output (TxD)
Pin 3: Data input (RxD)
Pin 4: Internal ground (GND)
Pin 5: Clear to send (CTS)
Pin 6: Not connected
Pin 7: Internal ground (GND)
Pin 8: Internal ground (GND)
Pin 9: Not connected
Pin 10: Not connected
Pin 11: Charging voltage for
rechargeable battery pack
+12 ... +20 V ( I _ out 25mA)
Pin 12: Reset _ Out *)
Pin 13: +5 V output
Pin 14: Internal ground (GND)
Pin 15: Universal remote switch
Pin 16: Not connected
Pin 17: Not connected
Pin 18: Not connected
Pin 19: Not connected
Pin 20: Data terminal ready (DTR)
Pin 21: Internal ground (GND)
Pin 22: Not connected
Pin 23: Not connected
Pin 24: Not connected
Pin 25: +5 V output
*) = Hardware restart
Preparation
You can set these parameters for other
devices in the Setup menu (see the chapter
entitled "Configuring the Balance"). You
will also find a detailed description of
the available data interface commands
in the file "Data Interface Descriptions
for AGB/PGB Models", which you can
download from the KERN website,
www.kern-sohn.com.
The many and versatile properties of these
14
1
balances can be fully utilized for printing
out records of the results when you connect
your balance to a KERN data printer. The
recording capability for printouts makes it
easy for you to work in compliance with
ISO/GLP.
For remote switch