Counting
Purpose
With the Counting program you can
determine the number of parts that each
have approximately equal weight.
Preparation
Configure the Counting application
in the operating menu:
See "Configuring the Balance"
Menu code: 2. 1. 4
Example
Determine an unknown piece count; weigh the preset reference sample quantity
Menu: Application program: Counting (menu code 2. 1. 4);
Reference sample quantity: 20 pcs (menu code 3. 3. 3)
Step
1. Tare the balance
2. Display the reference sample
quantity (here: 20 pcs)
3. Place the reference sample
quantity (20 pcs) on the balance
(here: 66 g)
4. Start the application; if the print
format is set to include data
ID codes, the following piece
weight is printed
5. Weigh uncounted parts
(here: 174 pcs)
6. Display weight
7. Display quantity
8. Unload the balance
9. Delete the reference value
10. Repeat the procedure starting from step 5, if desired
42
$ Reference sample quantity:
Code 3. 3. 1
Code 3. 3. 2
Code 3. 3. 3
Code 3. 3. 4
Code 3. 3. 5
$ Storage parameter
(display accuracy for counting)
Code 3. 4. 1
Code 3. 4. 2
See also "Configuring the Balance"
Key (or instruction)
t
v >2 sec.
v
v
v
c
5 pcs
10 pcs (factory setting)
20 pcs
50 pcs
100 pcs
Standard resolution
(factory setting)
With 10 times higher
resolution than standard
Display/Printout
0.0
g
ref 20
(briefly)
+
66.0
g
+
20
pcs
wRef
+
3.300 g
+
174
pcs
+
574.2
g
+
174
pcs
0
pcs