Weighing In Percent - KERN AGB Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Weighing in Percent

Purpose
This application program allows you to
obtain weight readouts in percent which
are in proportion to a reference weight.
Preparation
Configure the Weighing in Percent
application in the operating menu:
See "Configuring the Balance."
Menu code: 2. 1. 5
Example
Determine an unknown percentage: store the weight on the balance as a reference percentage
Menu: Application program: Weighing in Percent (menu code 2. 1. 5)
Menu: Reference percentage: 100 % (menu code 3. 3. 5)
Step
1. Tare the balance
2. Display the reference percentage:
3. Place the reference weight for 100 %
on the balance (here: 222.5 g)
4. Start the application; if the print
format is set to include data
ID codes the following is printed
5. Place an unknown weight on
the balance (here: 322.5 g)
6. Display weight
7. Display percentage
8. Unload the balance
9. Delete the reference percentage
10. Repeat the procedure starting from step 5, if desired.
$ Reference percentage:
Code 3. 3. 1
Code 3. 3. 2
Code 3. 3. 3
Code 3. 3. 4
Code 3. 3. 5
$ Storage parameter
(display accuracy for counting)
Code 3. 4. 1
Code 3. 4. 2
See also "Configuring the Balance"
Key (or instruction)
t
v >2 sec.
v
v
v
c
5 %
10 % (factory setting)
20 %
50 %
100 %
Standard resolution: With
stability (factory setting)
With 10 times higher
stability than standard
Display/Printout
0.0
g
ref100
+
222.5
g
+
100.00
%
Wxx%
+
222.500 g
+
144.94
%
+
322.5
g
+
144.94
%
0.00
%
43