I-Lift™
8.12 Checking and Tightening Mast Pivot Bolt
Check that the bolt A is through the bracket B and the locknut
1.
Cis tight and secure.
2.
If needed, do one or more of the following:
•
Tighten locknut and back-off the locknut 1/8 of a turn.
•
Replace the locknut.
46
8.13 Replacing the Swivel Bar
WARNING!
Risk of Injury
Worn or damaged parts of the lift may cause injury to
the patient or assistants.
– After the first year of use, the hooks of the hanger
bar and mounting brackets of the boom should be
inspected every six months to determine the extent
of wear. If these parts become worn, replacement
must be made.
If the scale is installed on the lift, refer to 9.3 Installing the
Scale, page 50 to replace the hanger bar.
1171892-B