Secondary Emergency Release
It is recommended that the primary emergency release be
used. The secondary emergency release is only a back-up to
the primary emergency release.
In cases where the primary release is either not functioning or
unreachable, a secondary emergency release may be used.
Pull up on the EMERGENCY grip A and push down on the
1.
boom B at the same time.
1171892-B
5.5 Performing an Emergency Stop
Press the RED button A on the control box B in to stop the
1.
boom and patient from raising or lowering.
2.
To reset, rotate the emergency button clockwise.
5.6 Mounting the Battery Charger
Refer to your local regulations concerning proper mounting
procedures.
Operation
23