Keysight Technologies RP7900 Serie Guide D'utilisation Et De Maintenance page 240

Masquer les pouces Voir aussi pour RP7900 Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5 SCPI Programming Reference
Parameter
(none)
Query the power line frequency: SYST:LFR?
SYSTem:LFRequency:MODE AUTO|MAN50|MAN60
SYSTem:LFRequency:MODE?
Specifies automatic or manual line frequency detection. 
AUTO specifies automatic detection.
MAN50 specifies a setting of 50 Hz.
MAN60 specifies a setting of 60 Hz.
This parameter is saved in non-volatile memory.
Parameter
AUTO|MAN50|MAN60
Sets the line frequency mode to 60 Hz: SYST:LFR:MODE MAN60
SYSTem:PASSword:FPANel:RESet
Resets the front panel lockout password to zero. This command does not reset the calibration
password.
Parameter
(none)
Resets the front panel password: SYST:PASS:FPAN:RES
SYSTem:REBoot
Reboots the instrument to its power-on state.
Parameter
(none)
Reboots the instrument: SYST:REB
SYSTem:SDS:CONNect
Connects the power supply to the Keysight SD1000A SDS unit. This command is used when the SDS
connect mode is set to MANual.
Parameter
(none)
Connect the SDS unit: SYST:SDS:CONN
240
Typical Return
50 or 60
Typical Return
AUTO, MAN50, or MAN60
Typical Return
<+0,"No error">
Typical Return
(none)
Typical Return
(none)