CALibrate Subsystem
Calibrate commands calibrate the instrument.
Read the
accuracy and reliability.
CALibrate:COUNt?
Returns the number of times the unit has been calibrated. The count is incremented whenever the
calibration (and date) is saved, the administration password is changed or reset, or the firmware is
updated.
Parameter
(none)
Return the calibration count: CAL:COUN?
CALibrate:CURRent[:LEVel] <value>
Calibrates the current programming and measurement. The value selects the range to calibrate.
Parameter
The maximum current of the output range.
Calibrates the current of the 10 A range: CAL:CURR 10
CALibrate:CURRent:SHARing
Calibrates the Imon signal for paralleled units.
Parameter
(none)
Calibrates the current sharing: CAL:CURR:SHAR
CALibrate:CURRent:TC
Calibrates the temperature coefficient.
Parameter
(none)
Calibrates the temperature coefficient: CAL:CURR:TC
Keysight RP7900 Series Operating and Service Guide
calibration section
before calibrating. Improper calibration reduces
5 SCPI Programming Reference
Typical Return
<count>
Typical Return
(none)
Typical Return
(none)
Typical Return
(none)
191