2 Installing the Instrument
This is a Delta-type with corner ground.
Configurations for 400/480 VAC units
Do not connect the Neutral to the instrument.
Surge Protector Installation
A Surge Protector (Kit p/n RP7961A-60007) is required for Keysight models RP7941A, RP7942A,
RP7943A, RP7945A, RP7946A, RP7961A, RP7962A, and RP7963A.
Equipment Damage. The steady-state line rating for the surge protector is 350 VAC
maximum line-to-earth; and 600 VAC maximum line-to-line.
Before getting started, check the following list and verify that you have received these items. If
anything is missing, please contact your nearest Keysight Sales and Support Office.
Line Surge Protector PCA 400V - p/n 5067-6846
Screw Torx-T10 M3X0.5 35 mm - p/n 0515-1038
Bolt Hex M5X0.8 35 mm qty 3 - p/n 0515-6395
SHOCK HAZARD. Disconnect the power cord from the AC mains before installing the
surge protector..
72
This is a Delta-type with high-leg ground.
This is a floating Delta-type.
Keysight RP7900 Series Operating and Service Guide