3 Getting Started

Set the Output Current

Method 1
Use the left and right navigation keys to navigate to the setting that you wish to change.
In the display below, the current setting is selected. Use the up and down navigation keys to toggle
between the + and - limit entries. Enter a value using the numeric keypad. Then press Select.
You can also use the numeric arrow keys to adjust the value up or down. You can set both positive and
negative current values. Values become effective when the output is turned on.
In current priority mode, the unit will maintain the output current at its programmed setting. In voltage
priority mode, the unit will limit the output current when it reaches the specified current limit value.
Refer to
Set the Output Mode
for more information.
Method 2
Use the Current key to select the current entry field. In the display below, the current setting is
selected. Enter the desired setting using the numeric keypad. Then press Enter.
If you make a mistake, either use the backspace key to delete the number, press Back to back out of
the menu, or press Meter to return to meter mode.
100
Keysight RP7900 Series Operating and Service Guide