INSTrument:GROup:PRIMary:CONNect:MODE AUTO|MANual
INSTrument:GROup:PRIMary:CONNect:MODE?
Specifies the connection mode of the primary unit. This setting is saved in non-volatile memory.
AUTO - the primary unit will try to connect to the previously discovered secondary units at power-on.
MANual - the primary will connect to the previously discovered secondary units when it receives a
connection command from the front panel or from INST:GROUP:PRIM:CONN.
Parameter
AUTO | MANual
Specifies automatic connection mode INST:GRO:PRIM:CONN:MODE AUTO
INSTrument:GROup:PRIMary:DISCover
Instructs the primary unit to discover all secondary units connected to the primary/secondary bus.
Parameter
(none)
Discover all secondary units INST:GRO:PRIM:DISC
INSTrument:GROup:PRIMary:RESet
Resets the primary-discovered secondary configuration. It disconnects any secondary-connected
units and returns them to independent operation.
Parameter
(none)
Resets all secondary units INST:GRO:PRIM:RES
INSTrument:GROup:SECondary:ADDRess <value>
Sets the secondary unit's bus address. Each secondary unit in a primary/secondary group must have a
unique bus address or bus communication will fail. This setting is saved in non-volatile memory.
Parameter
1 - 19
Set the secondary unit's address to 1 INST:GRO:SEC:ADDR 1
Keysight RP7900 Series Operating and Service Guide
5 SCPI Programming Reference
Typical Return
AUTO or MAN
Typical Return
(none)
Typical Return
(none)
Typical Return
1