Characteristic
Output on delay time 8
    Voltage priority - high impedance:
    Voltage priority - low impedance:
    Current priority:
Maximum load inductance: 9
Line regulation
    Voltage:
    Current:
Typical leakage current
    Output disabled - instantaneous: 10 :
    Output disabled - long term 11 :
    Unit unplugged from AC mains:
1 For lower and upper voltage compliance, respectively
2 For digitization at 5.12 µs sampling rate
3 Applies for ambient temperatures above 30°C or below 20°C
4 From occurrence of over-voltage to start of shutdown
5 In the CV Comp 0 range, Frequency = 100 kHz, under no load, and a step change from 10% to 100% of output rating
6 In the CV Comp 0 range, Frequency = 2.3 kHz, with full constant current load, and a step change from .1% to 100% of output rating
7 In the CC Comp 0 range, Frequency = 100 kHz, and a step change from 10% to 100% of current rating, into low impedance DC source
8 Time from when the Output On command is received to the initiation of turning the output on
9
Unit is susceptible to internal damage if maximum load inductance limits are exceeded at the rated current; refer to
load graphs
for additional information
10 Unit powered on with the output disabled for >1 second after full load operation
11 Unit powered on with the output disabled for >10 seconds after full load operation

Common Characteristics

Common Characteristic
Command Processing Time
Constant Dwell ARBs
    Number of points:
    Dwell range:
    Dwell resolution:
Computer Interfaces
    LXI:
    LAN:
    USB:
    GPIB:
Environmental
    Operating environment:
    Temperature range:
    Relative humidity:
    Altitude:
    Storage temperature:
Keysight RP7900 Series Operating and Service Guide
RP7931A,
RP7932A,
RP7941A
RP7942A
8.1 ms
8.1 ms
8.1 ms
8.1 ms
7.1 ms
7.1 ms
20 µH
200 µH
< 10 µV
< 10 µV
< 20 µA
< 10 µA
2 µA
800 µA
0.4 µA
800 µA
2 µA
800 µA
≤ 1 ms from receipt of command to start of output change. Applies to simple settings commands
over the GPIB interface (see
Typical Command Processing
Up to 65,535
One dwell setting applies for the entire ARB, from 10.24 µs to 0.30 seconds
Values are rounded to the nearest 10.24-microsecond increment
LXI 1.5 LXI Device Specification 2016
10 Mb, 100 Mb, 1 Gb LAN
USB 2.0 (USB-TMC488 protocol)
SCPI - 1993, IEEE 488.2 compliant interface
Indoor use, installation category II (for AC input), pollution degree 2
0°C to 55°C (Maximum continuous current available is derated at 1% of rating per degree C from
40°C to 55°C)
95% or less (non-condensing)
Up to 2000 meters
-30°C to 70°C
RP7933A,
RP7935A,
RP7943A
RP7945A
8.1 ms
8.1 ms
8.1 ms
8.1 ms
7.1 ms
7.1 ms
10 µH
100 µH
< 10 µV
< 10 µV
< 40 µA
< 10 µA
4 µA
800 µA
0.8 µA
800 µA
4 µA
800 µA
All Models
Times)
1 Quick Reference
RP7936A,
RP7946A
8.1 ms
8.1 ms
7.1 ms
400 µH
< 10 µV
< 10 µA
800 µA
800 µA
800 µA
inductive
37