Control And Display Devices; Status Symbols; Control Knob; Keys And Set-Up Options - Hettich ROTIXA 50 S Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12

Control and display devices

See figure on page 2.
Fig. 2:
Display and operating panel
12.1

Status symbols

Lid open.
Lid closed.
Rotation indicator. The indication appears while the centrifuge is running as long as the rotor is turning.
STOP
Centrifugation run stopped or finished. Indication after completion of the centrifugation run as long as the
rotor is turning. The indicator flashes following an emergency stop.
LOCK 1,
Keyswitch position (for centrifuges with keyswitch only).
LOCK 2
LOCK 4,
Program interlocking when serial communication provided (for centrifuges with serial communication only).
LOCK 5
PC, PC
Serial communication (for centrifuges with serial communication only).
Operating errors or faults that occur are shown symbolically in the display (see Chapter "Faults").
12.2

Control knob

For setting the individual parameters.
Turning anticlockwise reduces the value. Turning clockwise increases the value.
12.3

Keys and set-up options

Running time, parameter t/min:sec
t
1.
Parameter t/min: Adjustable from 1 - 999 min, in 1 min steps.
2.
Parameter t/ :sec Adjustable from 1 - 59 s, in 1 second steps.
3.
Continuous running "---:--"
Interrogation of the integral RCF, parameter ∫RCF.
RCF
1. Speed, parameter RPM
n
A numerical value from 50 RPM up to the maximum speed of the rotor (n-max-Rotor) can be set. Maximum
speed of the rotor, see Chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors and accessories".
Adjustable in steps of 10.
2. Interrogation of the maximum speed of the rotor, parameter n-max-Rotor.
1. Relative centrifugal force, parameter RCF/RZB
RCF
A numerical value can be set, which gives a speed between 50 RPM and the maximum speed of the rotor
(n-max-Rotor). Adjustable in steps of 1.
2. Interrogation of the maximum RCF of the rotor, parameter RCF-max-Rotor.
Run-up parameters
1. Run-up steps, parameter
2. Run-up time, parameter
Run-down parameters
1. Braking steps, parameter
Step R9, B9 = short run-down time, ... Step R1, B1 = long run-down time, Step R0 = unbraked run-down.
2. Run-down time, parameter
3. Brake de-energisation speed, parameter n
After reaching this speed, unbraked run-down occurs.
1. Temperature setpoint, parameter T/°C (for centrifuges with cooling only). Adjustable from -20°C to +40°C,
in 1°C steps (heating/cooling option settable between -20°C and +60°C / +90°C). The lowest temperature
that can be achieved is dependent upon the rotor (see Chapter "Anhang/Appendix, Rotoren und
Zubehör/Rotors and accessories").
2. Centrifuging radius, parameter r/mm. Entry in mm. Centrifuging radius, see Chapter "Anhang/Appendix,
Rotoren und Zubehör/Rotors and accessories".
. Step 9 = shortest run-up time, ... Step 1 = longest run-up time.
1-9
. The time range that can be set is dependent upon the set speed.
min:sec
t
. R = linear braking curve, B = similar to an exponential braking curve.
0-9
. The time range that can be set is dependent upon the set speed.
t
min:sec
(•)
/RPM
EN
33/104

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Rotixa 50 rs

Table des Matières