Medela Symphony Mode D'emploi page 168

Masquer les pouces Voir aussi pour Symphony:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

13. EMC / Technical description
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
a
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the electric breastpump Symphony is used exceeds the
applicable RF compliance level above, the electric breastpump Symphony should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the electric breastpump Symphony.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
b
168

Publicité

Table des Matières
loading

Table des Matières