Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
Electromagnetic immunity
The electric breastpump Symphony is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Symphony should assure
that it is used in such an environment.
IEC 60601
Immunity test
test level
3 Vrms
Conducted RF
150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
Compliance
Electromagnetic
level
environment – guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the electric
breastpump Symphony, including
cables, than the recommended se-
paration distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 0,35√P
10 Vrms
d = 1,2√P
3 V/m
d = 2,3√P
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manu-
facturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmit-
ters, as determined by an electroma-
gnetic site survey
than the compliance level in each
frequency range.
Interference may occur in the vicinity
of equipment marked with the fol-
lowing symbol:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
, should be less
a
b
167