6.6.2 Leica EL6000 compact light source
• Instructions on inserting the lamp are given in
the Leica EL6000 compact light source
manual.
• The microscope adapter for connecting the
light guide is attached to the back of the stand
(22.1).
Caution!
Connect the light guide to the microscope
first to prevent exposing the user to the
high-energy light output of the Leica EL6000
compact light source.
• Insert the light guide (27.1) into the
microscope adapter (27.2) and secure it with
the clamping screw. Insert the opposite input
of the light guide into the port (28.1) on the
rear panel of the compact light source. The
light guide must click into place.
!
Caution!
Take care not to kink or otherwise damage the
light guide when connecting it to the light
source
or
microscope
overtighten the clamping screw.
Use only light guides with Storz long connectors
to prevent damage to the unit and danger to the
user (blinding hazard).
adapter.
Do
not
Caution!
Connect the light guide at both ends (light
source/adapter) before opening the shutter
or attenuator.
The emitted light may result in eye or skin
injuries or damage to materials.
Never look directly into the light emitted by
the light guide.
Further information → Leica EL6000 manual.
Fig. 27
Light guide with adapter
1
Light guide
2
Adapter for Leica microscopes
1
Fig. 28
Rear panel with connectors
1
Light port
2
Remote control port
3
AC input
1
3
6. Assembly
2
2
31

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Dm6000 bDm6000 m

Table des Matières