5. For your safety / 6. Packaging
Remove the pressure transmitter from service and mark it to prevent it from
!
being used again accidentally, if it becomes damaged or unsafe for opera-
tion
Warning
Take precautions with regard to remaining media in removed pressure
transmitter. Remaining media in the pressure port may be hazardous or
toxic!
Have repairs performed by the manufacturer only.
The operator is responsible for the material compatibility as well as correct
handling, operation and maintenance.
Information about material consistency against corrosion and diffusion can be found in our
WIKA-Handbook, 'Pressure and Temperature Measurement'.
Consider the details given in the EC-type examination certificate as well
as the respective country specific regulations for installation and operation
in hazardous areas (e.g.: IEC 60079-14, NEC). If you do not observe these
Warning
stipulations, serious injuries and/or damage can occur.

6. Packaging

Has everything been supplied?
Check the scope of supply:
Completely assembled transducer
Inspect the transducer for possible damage during transportation. Should there be
any obvious damage, inform the transport company and WIKA without delay.
The transducer is purified, evacuated and double-bagged for clean-room service
and should remain packaged until installation. A protective cap is used to protect the
face sealing beads from nicks or scratches. This plug should remain in place during
storage and handling, in order to prevent damage to the bead. Therefore remove the
inner ESD (Electro-Static Discharge) bag only at the place of installation.
Keep the packaging, as it offers optimal protection during transportation (e.g. chan-
ging installation location, shipment for repair).
6
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio NWU-1
GB
6. Packaging / 7. Starting, operation
Ensure that the pressure connection thread and the connection contacts will not be
damaged.
Remove this protection cap only just before installing the pressure transmitter
Mount the protection cap when removing and transporting the instrument.
Unpack the transducer
1. Remove the transducer from the box.
2. Remove the outer protective bag and discard.
3. Carry the transducer (sealed in the inner bag), into the clean area.

7. Starting, operation

Required tools: wrench (flats 19 and 16), screw driver (0.040" to 0.060" /
1 to 1.5 mm) and a pair of scissors, allen key for WU-16
Mechanical connection
Don't use this
spanner flat!
Use this spanner flat
for screwing in!
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio NWU-1
S
Product label (example)
Signal
Power Supply
P #
Product No.
S #
Serial No.
PIN assignment
Coded manufacture date
GB
7
S