FHR Performance :
Complies With:
Classification:
Environmental:
A full technical description is provided in the Service Manual 752333.
Attention, consult accompanying
documents / Instructions to Use
TECHNICAL SPECIFICATION
Standard Mode
Range - 60-210bpm
Resolution - 1bpm
Averaging - 4 beats
Accuracy - ±3bpm
IEC 60601-1:1988/A2:1995, EN 60601-1:1990/A13:1996, UL 60601-1
1st Edition, CAN/CSA C22.2 No. 601.1-M90
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY, IN ACCORDANCE WITH UL60601-1 and CAN/CSA
C22.2 No 601.1.
Type of shock protection
Degree of shock protection
Protection against water ingress
Degree of safety in presence of
flammable gases
Mode of operation
Operation
Temperature
Relative Humidity
Pressure
Transport and Storage
Temperature
Relative Humidity
Pressure
Medical Devices Directive 93/42/EEC
8
Smoothed Mode
Range - 60-210bpm
Resolution - 1bpm
Averaging - 8 beats
Accuracy - ±3bpm
Class 1 mains powered equipment
Type B equipment
Not Protected - IPX0
Equipment not suitable for use in
presence of flammable gases
Continuous
:
+10°C to +30°C
:
30% to 75% (non condensing)
:
700 to 1060hPa
:
-10°C to +40°C
:
20% to 90% (non condensing)
:
700 to 1060hPa
Attention: consult this manual.
Refer to safety section.
Manual Mode
Range - 60-210bpm
Resolution - 1bpm
Averaging - 10 beats
Accuracy -±3bpm
(excluding user error)