22 ArthroCare
Atlas™ System
®
User's Manual
Maintenance and Troubleshooting
Maintenance
Other than fuse replacement, the Controller has no user-serviceable parts. It is designed to provide consistent out-
put levels and is calibrated by clock crystals, voltage references, and fixed resistors. There are NO internal adjust-
ments in the instrument and, due to the integrated calibration methods, no annual maintenance check is required.
There is NO software incorporated in the ArthroCare Atlas Controller. If any component malfunctions, call Customer
Service for a return authorization.
Fuse Replacement
The fuse holder is located on the back of the Controller. To replace a fuse, turn off the power to the Controller and
unplug the Power Cord from the power outlet at the rear of the Controller. After waiting at least 10 seconds for inter-
nal circuitry to discharge, use a screwdriver or similar tool to remove the fuse holder by depressing the locking tabs.
Replace both fuses with the same type and rating as specified on the rear panel of the Controller. Reinsert the fuse
holder until the locking tabs snap into place. Reconnect the Power Cord and restore power to the Controller. If a
fuse fails again, disconnect all power to the Controller and contact Customer Service.
Troubleshooting Guide
If you are experiencing problems with the ArthroCare Atlas System, you may want to use the following troubleshoot-
ing guide to help identify or eliminate the problem before contacting Customer Service:
• System does not power up after the power switch is pressed
Check that the Power Cord is properly connected to the Controller and plugged into an appropriately grounded
outlet. If the unit is plugged in properly, check if the fuses have blown. To change the fuses, follow the instruc-
tions for Fuse Replacement.
• Green Foot Control / Hand Control Connected Indicator light does not illuminate
Check that the Foot Control / Hand Control is properly connected to the Controller and that the cord is not nicked,
cut or frayed. Do not use the Foot Control / Hand Control if the cord has been damaged. If the problem persists,
change the Foot or Hand Control. If the Foot Control/Hand Control Connected Indicator Light is still not illuminat-
ed, return the System for Service.
• Green Wand Connected Indicator light does not illuminate when a Wand is connected to the reusable
Patient Cable.
Make sure that the Wand is securely seated in the Patient Cable, and that the Patient Cable is properly connected
to the Controller. If the problem persists, first change the Patient Cable and then the Wand. If the Wand indicator
light is still not illuminated, return the System for service.