Guidance And Manufacturer's Declaration - Electromagnetic Emissions - smith&nephew SPIDER2 Limb Positioner Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The SPIDER2 is intended for use in the electromagnetic environment specified below. The customer or the user of the SPIDER2 should ensure that it is used in such an
environment.
Emission Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/flicker
Not applicable
emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Guidance for Separation Distances
The SPIDER2 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the SPIDER2 can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the SPIDER2 as
recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output Power of
Transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
SPIDER2 Limb Positioner
Instructions for Use
Electromagnetic Environment – Guidance
The SPIDER2 uses RF energy only for its internal functions. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The SPIDER2 is suitable for use in all establishments, including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for domestic purposes.
Recommended Separation Distances Between Portable and Mobile
RF Communications Equipment and the SPIDER2
150 KHz to 80 MHz
d = 1.7 √ P
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Separation Distance According to Frequency of Transmitter M
80 MHz to 800 MHz
d = 1.7 √ P
0.12
0.37
1.17
3.69
11.67
English
800 MHz to 2.5 GHz
d = 2.34 √ P
0.23
0.74
2.34
7.38
23.34
126-99-01 Rev. C