Cleaning and Sterilization (Cont.)
Gravity Displacement Steam Sterilization
The following parameters are recommended for gravity displacement steam sterilization:
• Set temperature at 270-272°F (132-133°C);
• Set exposure time to 15 minutes for wrapped Cables;
• Set exposure time to 10 minutes for unwrapped cables;
• Set drying time for 8 minutes minimum.
Or
• Set temperature at 250-254°F (121-123°C);
• Set exposure time to 35 minutes for wrapped or unwrapped cables;
• Set drying time for 8 minutes.
CAUTION:
Failure to properly clean the Patient Cable may lead to inadequate sterilization.
CAUTION:
The recommended sterilization parameters are only valid with sterilization equipment that is properly
maintained and calibrated.
CAUTION:
To insure adequate sterilization, make sure that the protective caps do not cover the cable connectors
during sterilization.
CAUTION:
Make sure Cable is thoroughly dry before use. Wet Cable may damage the Cable and Controller.