26 Maintenance
This appliance requires some maintenance to meet its requirements and maintain its general
characteristics.
BATTERY
STRAPS
Back Light SCREEN
Measurement setting /
check
CONNECTIONS
This "advice" does not engage the responsibility of ALCAD.
It guarantees the best possible use of the characteristics and the preservation of the product.
Routine maintenance:
The basic maintenance is simply cleaning the outside of the appliance. Any other operation requires a
trained personal.
Unplug the appliance before any intervention.
Do not let water flow inside the appliance: risk of electric shock.
Regularly clean the appliance under the following conditions:
use soapy water
-
never use any product containing petrol, benzene, alcohols that would attack silkscreen printings
-
wipe out with a soft lint-free cloth
-
use a solvent-free antistatic product to clean the screen.
-
For the carter:
clean it with a clean dry cloth
-
no solvent allowed.
-
98
Consequences
Recommended periodicity of
controls
Reduction of the battery
life
Breakdown
At each use
Check the holding of the straps
Reduction of visibility
Erroneous measures
Once a year
Erroneous measures
At any measurement
Recommended use
limit
200 charge /
discharge cycles
or 2 years
2 years
18 months