2. RECOmmENDATIONS FOR ELECTRIC CONNECTION
- Use cables with an operating limit temperature of at least 80 °C.
- For normal operating conditions the measuring signal can be transmitted by a shielded
cable of 0,75 mm
2
- The cable must not be installed in combination with carrying lines with a higher voltage
or frequency.
- If a combined installation cannot be avoided, a minimum space of 30 cm should be
respected.
- The power supply must be filtered and regulated.
2.3.1 assembling cable plug eN175301-803
-
Extract part [3] from part [2]
-
Unscrew cable gland [5]
-
Insert cable into part [2] through cable gland [5]
-
Wire part [3] as described in
paragraph 2.3.2, 2.3.4 or 2.3.5
-
Insert part [3] back
-
Fasten cable gland [5]
-
Place gasket [4] between the cable plug and the fixed connector
-
Connect the cable plug to the sensor
-
Fasten screw [1] to ensure tightness and correct electrical contact.
2.3.2 Wiring the hall version with 2 pulse outputs
This version has one NPN transistor output and one PNP transistor output.
V+ (12-36 VDC)
Max. wiring distance is 50 metres.
8020
INSTALLATION
cross section.
PNP transistor output
1
3
0 V
NPN transistor output
2
3
2
4
1
5