TC-TM 66XX
F. SOFTWARE UPDATE
The software is updated by the UPG32 program available on CD-ROM supplied with the product. To find out which
version of firmware is installed in your unit, use the
The quickest way to find out if an update is available is to visit the AOIP S.A.S website and look at the "Software" page.
To update the firmware, proceed as follows:
1.
If necessary, install on the PC the USB driver for communication with AOIP instruments. This driver can be
downloaded from our site, along with an information page describing the installation procedure
2.
Disconnect the leads connected to the measurement and simulation terminals.
3.
Connect the instrument to the PC using the USB lead.
4.
Download and run the firmware update program.
5.
Select the language then the file containing the firmware and download in the first stage.
6.
Choose the communication parameters that match the parameters of the TC6621. The communication port used
is a virtual port which does not correspond to a physical port on your computer. The other parameters to be
selected are defined in the diagram below.
Confirm the update by pressing "OK" and wait for the firmware to load into the unit.
7.
92 / 101