D.5.2
Configuration of the ramp generation
The CONFIGURATION/RAMP menu is used for the generation of incremental, single or cyclic ramp signals.
Incremental ramp signal configuration?
The figure below illustrates the type of single ramp that can be generated and their parameters:
Programming parameters in Increases mode
INC
T0
Tdt
To : time pressing the generation start key
Td : Time frame
Tdt : Duration
NB : Low Level
NH : High Level
INC: Increment (value of the increments as a %, in mVolt, °C or °F)
The LOW level and HIGH level levels are expressed:
As a percentage of the range if the scaling mode is ON.
In mVolt or in temperature units if the scaling mode is OFF and according to the type of value emitted (voltage or
temperature emission).
The Duration corresponds to the incrementation time it takes to go from Low Level to High Level (and vice versa with
decrementation). It is given in seconds and is limited to a maximum of 1000s.
The Delay corresponds to the timeout that can be programmed between the moment the start of transmission key is
pressed and the actual start of generation. It is given in seconds and is limited to a maximum of 1000s.
Use the F2 key to move to the next field.
Use the navigation keys to enter the value:
As a percentage of the gauge if the scaling mode is ON.
and to increase or decrease the value
And to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press VAL to save the parameters.
To quit the menu without saving, press CLEAR.
Storage of acquisitions in progress
NH
NB
Td
T0
Tdt
Td
T0
Tdt