TC-TM 66XX
C.2
Voltage / resistance or temperature simulation
To access the Emission mode:
The choice of Emission mode is made using the F2 key (mode menu).
Using the navigation keys ( and ), position the cursor in the Emission field going down the menu.
Confirm your choice using the VAL key.
Once you have confirmed the Emission mode, define the type of generation:
Continuous (manual or arrows or default editing).
Incremental (by step or "manual default").
By single ramp (only one ramp emitted).
By cyclic ramp.
Synthesizer ("automatic default").
Voltage / resistance or temperature generation / manual editing?
Press the F2 key to display the edit menu.
Using the navigation keys ( and ), choose the Manual edit mode and confirm (VAL key).
Press VAL again and enter your value using the navigation keys:
and to increase or decrease the value
o
o
and to select the digit to modify (hundreds/tens/unit/decimal place/hundreds).
Beware: if the Scaling mode is ON, the value to edit is expressed in the unit defined, otherwise this value is expressed in
Volts or in °C or °F.
70 / 101
NTA47230-300A6
PROGRAMMING MODES