D.4
Configuration of default value points
The configuration of default value points is performed in the configuration/Points menu, obviously providing the default
values mode has been confirmed.
Using the F1 key, select the configuration/Points menu.
Confirm using the VAL key.
Using the F1 key (
Using the navigation keys ( and  ), select the digit that requires editing.
Using the navigation keys ( and ), increase the value of the selected digit.
Confirm the line using the VAL key.
You can edit a value already recorded:
Using the navigation keys ( and ), select the line requiring editing.
Press the navigation key () to make editing possible.
You can delete a value already recorded:
Using the navigation keys ( and ), select the line requiring deleting.
Press the
key to delete the line.
Notes:
A maximum of 100 values can be entered.
This table of values is also used for the Synthesizer mode; consequently all changes to this table entail a change to the
synthesizer values.
) add a new value line to the table.