Materials used

- Nebuliser handset: Polypropylene, ther-
moplastic elastomer, polyoxymethylene
- Mouthpiece: Polypropylene, silicone
rubber
®
Lamira
contains no parts made from
natural rubber (latex).
Ambient conditions for
operation
- Ambient temperature:
+5 °C to +40 °C
- Relative humidity of ambient air:
15% to 93% (non-condensing)
- Air pressure: 700 hPa to 1060 hPa
Within professional healthcare facility
environments, the device must only be
used in in-patient wards and intensive
care units. The device must never be used
in areas exposed to elevated magnetic or
electrical radiation (e.g., in the vicinity of
an MRI scanner). The device must not be
used in closed proximity to HF surgical
equipment.
Users can operate the device both at
home and when out and about. Its use on
trains and aeroplanes is restricted to pas-
senger areas only. If using the device in a
car, it must be operated with battery
power.
CAUTION:
When using portable wireless communi-
cation devices, you must ensure that a
distance of at least 30 cm is maintained
between all parts of the nebuliser system
and these devices (including their acces-
sories, e.g. antenna cables or external
antennas). Otherwise, the performance of
your nebuliser system may deteriorate.
54
CAUTION:
The controller must not be placed directly
beside or stacked with other devices for
operation. If operation near or stacked
with other devices is unavoidable, the
controller must be monitored continuously
to ensure proper operation.
Storage and transportation
conditions
- Temperature: -25 °C to +70 °C
- Relative humidity of ambient air:
0% to 93% (non-condensing)
- Air pressure: 500 hPa to 1060 hPa
NOTE:
Moisture from condensation can impair
the controller's functional capabilities.
Extreme temperature fluctuations should
therefore be avoided. Wait until the con-
troller has adjusted to room temperature
before you start an inhalation session.
If the device is not used for a prolonged
period, the (rechargeable) batteries
should be removed first.
Info:
Batteries vary widely with regard to
life cycle and performance depending
on their quality. The operating time of
about 90 minutes can only be
achieved with the batteries (4x)
included with the delivery.
®
Lamira
- 2020-08