4 First use

WARNING
Hazard from nonsterile products .
Infection danger to the care provider and patient .
 Before first use and after each use, prepare and sterilise the medical device if needed .
CAUTION
Damage from soiled and moist cooling air .
Contaminated and moist cooling air can cause malfunctions and lead to premature bearing wear .
 Make sure that the supply of cooling air is dry, clean and uncontaminated according to ISO 7494-2 .
4.1 Check the amount of water
CAUTION
Overheating of the tooth due to lack of cooling water. Thermal damage to the dental pulp .
 Adjust the water amount for the spray cooling to a minimum of 50 cm3/min!
CAUTION
Hazard from insufficient spray water .
Insufficient spray water can cause the medical device to overheat and damage the tooth .
 Check the spray water channels and clean the spray nozzles with the nozzle needle if necessary .
MC01 B / BL / G / GL - Instructions for use - EN
37