MC01 B / BL / G / GL - Instructions for use - EN
Lightly spray O-rings on motor coupling with Faro Spray.
Attach the medical device to the motor coupling and turn it until the guide stud audibly snaps into place.
Pull on the medical device to make sure that it is securely affixed to the coupling.
5.2 Remove the medical device
Unlock the medical device from the motor coupling by twisting it slightly and then pulling it along its axis.
5.3 Inserting the milling tool or diamond grinder
Only use carbide cutters or diamond burs that correspond to ISO 1797-1type 1, are made of steel or tungsten carbide and meet the following criteria:
- shaft diameter: 2.334 to 2.35 mm
- overall length: max. 22 mm
39