MC01 B / BL / G / GL - Instructions for use - EN
Depending on the available autoclave, select a suitable procedure from the following sterilisation processes:
▪ Autoclave with three times initial vacuum:
– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4°F)
▪ Autoclave using the gravitation method:
– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4°F) or alternatively
– at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4°F)
Use according to the manufacturer's Instructions for Use.
7.8 Storage
Prepared products must be stored, protected from germs (as far as possible) and dust, in a dry, dark, cool room.
Comply with the expiry date of the sterilised items.
8 Material summary
Available from dental suppliers
Faro Spray
Nozzle needle