Faro MC01 B Mode D'emploi page 47

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MC01 B / BL / G / GL - Instructions for use - EN
7.2.1 Cleaning: Manual cleaning – external
Accessories required:
▪ Tap water 30 °C ± 5 °C (86 °F ± 10 °F)
▪ Brush, e.g. medium-hard toothbrush
 Brush off under flowing tap water.
7.2.2 Cleaning: Automated external cleaning
Faro recommends thermodesinfectors in accordance with EN ISO15883-1 that are operated with alkaline cleaning agents at a pH value of max. 10
 For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector (complying with max. pH value of 10).
 In order to prevent negative effects on the medical device, make sure that the interior and the exterior of the medical device are dry, and then lubricate immediately with
care agents Faro spray.
7.2.3 Cleaning: Manual cleaning – internal
Only automated interior cleaning with thermodisinfectorsin accordance with ISO 15883-1.
7.2.4 Cleaning: Automated internal cleaning
Faro recommends thermodesinfectors in accordance with EN ISO15883-1 that are operated with alkaline cleaning agents at a pH value of max. 10
 For program settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the thermodisinfector (complying with max. pH value of 10).
 In order to prevent negative effects on the medical device, make sure that the interior and the exterior of the medical device are dry, and then lubricate immediately with
care agents Faro spray
45

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mc01 blMc01 gMc01 gl

Table des Matières