Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MC01 B / BL / G / GL - Instructions for use - EN

5 Operation

5.1 Attach the medical device
WARNING
Detachment of the medical device during treatment.
A medical device that is not properly locked in place can become disconnected from the motor coupling and fall off.
 Carefully pull on the medical device before each treatment to ensure that it is securely locked onto the motor coupling.
CAUTION
Removing and attaching the contra-angle handpiece while the drive motor is rotating.
Damage to the drive.
 Never attach or remove the contra-angle handpiece while the drive motor is rotating!
CAUTION
Damage.
If the footswitch is pressed while you are connecting and removing the medical device, you can damage the medical device and motor coupling.
 Do not connect or remove the medical device while pressing the footswitch.
38

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mc01 blMc01 gMc01 gl

Table des Matières