6.2.1 Troubleshooting: Exchanging the O-rings on the motor coupling
CAUTION
Hazard from improper care of the O-rings.
Malfunctions or complete failure of the product.
Do not use Vaseline or other grease or oil.
NOTE
The O-rings on the motor coupling may only be lubricated with a cottonball wetted with Faro spray.
Press the O-ring between your fingers to form a loop.
Shove the O-ring to the front, and remove it.
Insert new O-rings into the grooves.
6.2.2 Troubleshooting: Cleaning the spray nozzle
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage the tooth.
Check spray water channels and if necessary clean spray nozzles with the nozzle needle.
Clean the water passage in the spray nozzles by using the nozzle needle.
MC01 B / BL / G / GL - Instructions for use - EN
43