Static Alignment; Dynamic Alignment - Össur TOTAL KNEE 2100 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
Patient Weight
50 - 70 kg (110-155 lbs)
70 - 90 kg (155-200 lbs)
90 - 110 kg (200-240 lbs)
110 - 125 kg (240-275 lbs)
• Remove bumper (leave the shims under it in place).
• Install appropriate bumper according to patient weight.
• Re-install the bumper cover by hooking the groove under the chassis
front and pushing the two pins into the chassis holes.
• Active patients may require a stiffer bumper
• Patients who have previously worn a prosthesis without a stance-
flexion feature may initially prefer a stiffer bumper.
• We suggest trying a softer bumper after two weeks of trial. This helps
the patient develop a security with the knee. Doing this will also help
make the transition to the stance-flexion characteristic more
acceptable.

STATIC ALIGNMENT

For safety, please make initial adjustments with patient standing between
parallel bars!
• Fit prosthesis and check for correct length.
• Make sure foot is flat on the floor, knee is in neutral position and
socket angles are correct.
• When the patient shifts weight onto the prosthesis, the geometric
lock is activated and the bumper should be slightly compressed.
Modify alignment if necessary.
• Explain function of stance flexion/geometric lock to patient. Ask
patient to activate it by stepping forward with prosthesis and
transferring weight onto heel.
• Patient should become accustomed to activating geometric lock and
resulting stance flexion movement.
• Knee cannot collapse as long as geometric lock is activated.
• Preset swing flexion resistance, Valve F (15, Figure 9)
a. Sit patient on chair. Extend prosthetic knee. Let it fall into flexion.
b. Increase resistance of Valve F until a small bump observed at 60°
flexion angle.

DYNAMIC ALIGNMENT

The model 2100 knee provides smooth and fluid swing control during
normal ambulation and allows changes in cadence.
For safety, please make initial adjustments with patient standing between
parallel bars!
Valve adjustment – F, H, E (Figure 9)
Valve
Factory setting
Valve F
3/4 open
Valve H
3/4 open
Valve E
1 1/2 open
Use small flathead screwdriver for adjustment.
Colour
Black
Red
Yellow
Orange
Function
Affects swing flexion resistance from 60°-160°
(fast walking only)
Affects swing flexion resistance from 0°-60°
(slower walking)
Affects whole swing extension resistance, 160°-0°
Type
Soft
Regular
Firm
Extra Firm
8

Publicité

Table des Matières
loading

Table des Matières